The study reveals that the hybrid closed loop device, MiniMed 670g, helps in reducing A1c by 0.75% compared to standard insulin pumps. For those transitioning to it, carbohydrate to insulin ratio needs to be adjusted immediately. This semi-automated system, differently called semi-artificial pancreas, is a significant advancement in the field of type 1 diabetes. The device was approved by FDA in September 2016.
GT
Also see

Diabetes Care
670G pivotal trial
April 2018
OBJECTIVE
The MiniMed 670G System is the first commercial hybrid closed-loop (HCL) system for management of type 1 diabetes. Using data from adolescent and young adult participants, we compared insulin delivery patterns and time-in-range metrics in HCL (Auto Mode) and open loop (OL). System alerts, usage profiles, and operational parameters were examined to provide suggestions for optimal clinical use of the system.
METHODS
Data from 31 adolescent and young adult participants (14–26 years old) at three clinical sites in the 670G pivotal trial were analyzed. Participants had a 2-week run-in period in OL, followed by a 3-month in-home study phase with HCL functionality enabled. Data were compared between baseline OL and HCL use after 1 week, 1 month, 2 months, and 3 months.
RESULTS
Carbohydrate-to-insulin (C-to-I) ratios were more aggressive for all meals with HCL compared with baseline OL. Total daily insulin dose and basal-to-bolus ratio did not change during the trial. Time in range increased 14% with use of Auto Mode after 3 months (P < 0.001), and HbA1c decreased 0.75%. Auto Mode exits were primarily due to sensor/insulin delivery alerts and hyperglycemia. The percentage of time in Auto Mode gradually declined from 87%, with a final use rate of 72% (−15%).
CONCLUSIONS
In transitioning young patients to the 670G system, providers should anticipate immediate C-to-I ratio adjustments while also assessing active insulin time. Users should anticipate occasional Auto Mode exits, which can be reduced by following system instructions and reliably bolusing for meals. Unique 670G system functionality requires ongoing clinical guidance and education from providers.